PTW729和ArcCheck探测器在调强放射治疗剂量验证中的应用 被引量:13

Application of PTW729 and ArcCheck detector in dose verification of intensity modulated radiotherapy
在线阅读 下载PDF
导出
摘要 目的:探索在适形调强放疗(IMRT)剂量验证中应用三维探测器ArcCheck替代二维探测器PTW729的可行性和优越性,分析影响PTW729、ArcCheck进行IMRT计划剂量验证通过率的主要因素。方法:分别使用ArcCheck和PTW729对IMRT计划进行验证测量,基于不同标准参数进行Gamma分析,采用两两比较的配对t检验对相同标准参数下的两种验证工具的γ通过率进行统计学分析,再利用非线性拟合的方法分析不同标准参数对两种工具验证结果的影响。结果:在临床常用标准参数下(剂量偏差>2%,距离偏差>2 mm,TH10),ArcCheck与PTW729对IMRT计划的剂量验证结果差异无统计学意义(P>0.05);影响上述两种验证工具γ通过率的主要因素分别是剂量偏差和距离偏差。结论:在临床常用标准参数范围内,三维验证系统ArcCheck是一种理想的IMRT剂量验证工具,其验证结果能更全面准确地反映实际治疗的各种信息。 Objective To explore the feasibility and superiority of applying three-dimensional detector ArcCheck to replace two-dimensional detector PTW729 in the dose verification of intensity-modulated radiotherapy; to analyze the main factors influencing the Gamma pass rates of intensity modulated radiotherapy(IMRT) plans based on PTW729 and ArcCheck.Methods ArcCheck and PTW729 were respectively applied to verify and measure the IMRT plans. Based on the Gamma analysis with different standard parameters, the Gamma pass rates of the two verification methods with the same standard parameters were compared by using the paired t-test. The non-linear fitting method was used to discuss on the effects of different standard parameters on the verification results of two methods. Results Based on the usual clinical standard parameters(dose deviation is greater than 2%; distance deviation is greater than 2 mm; threshold parameter is 10), the verification results of PTW729 and ArcCheck didn't show any statistical differences(P0.05). The dose deviation and distance deviation were the main factors influencing Gamma pass rates of IMRT plans. Conclusion Within the range of usual clinical parameters, three-dimensional system ArcCheck is an ideal tool for the dose verification of IMRT plans, and the results reflect the information of actual treatment more comprehensively and accurately.
出处 《中国医学物理学杂志》 CSCD 2016年第5期473-477,共5页 Chinese Journal of Medical Physics
基金 安徽省高校省级自然科学研究项目(KJ2010B380)
关键词 强调放射治疗 剂量验证 ArcCheck PTW729 Gamma分析法 标准参数 intensity modulated radiotherapy dose verification ArcCheck PTW729 Gamma analysis standard parameter
  • 相关文献

参考文献11

  • 1EZZELL G A, GALVIN J M, LOW D, et al. Guidance document on delivery, treatment planning and clinical implementation of IMRT: report of the IMRT subcommittee of the AAPM radiation therapy committee[J]. Med Phys, 2003, 30(8): 2089-2115.
  • 2CHANDARAJ V, STATHAKIS S, MANICKAM R, et al. Comparison of four commercial devices for RapidArc and sliding window IMRT QA[J]. JAppl Clin Med Phys, 2011, 12: 338-349.
  • 3POPPE B, BLECHSCHMIDT A, DJOUGUELA A, et al. Two- dimensional ionization chamber arrays for IMRT plan verification [J]. Med Phys, 2006, 33: 1005-1015.
  • 4LETOURNEAU D, PUBLICOVER J, KOZELKA J, et al. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy[J]. Med Phys, 2009, 36(5): 1813-1821.
  • 5LOW D A, HARMS W B, MUTIC S, et al. A technique for the quantitative evaluation of dose distributions [J]. Med Phys, 1998, 25(5): 656-661.
  • 6张伟,马照,邵鹏,马金波,刘培继,刘鹏.Delta4的临床应用研究[J].中华放射肿瘤学杂志,2010,19(1):48-51. 被引量:11
  • 7KOZELKA J, ROBINSON J, NELMS B, et al. Optimizing the accuracy of a helical diode array dosimeter: a comprehensive calibration methodology coupled with a novel virtual inclinometer [J]. Med Phys, 2011, 38(9): 5021-5032.
  • 8FEYGELMAN V, ZHANG G, STEVENS C, el al. Evaluation of a new VMAT QA device, or the "X" and "O" array geometries[J]. J Appl Clin Med Phys, 2011, 12(2): 146-168.
  • 9PETOUKHOVA A, EGMOND J, EENINK M, et al. The ArcCheck diode array for dosimetric verification of HybridArc [J]. Phys Med Biol, 2011, 56: 5411-5428.
  • 10LI G J, ZHANG Y J, JIANG X Q, et al. Evaluation of the ArcCheck QA system for IMRT and VMAT verification [J]. Phys Med, 2013, 29: 295-303.

二级参考文献10

共引文献26

同被引文献91

引证文献13

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部