期刊文献+

除颤仪性能参数的质量管控检测探讨 被引量:3

Discussion on Quality Control and Detection of Defibrillator Performance Parameters
在线阅读 下载PDF
导出
摘要 为了探讨除颤仪性能参数的质量管控措施,本文以某三级甲等医院6个品牌、9个型号的40台除颤仪作为监测对象,并对其性能参数检测结果进行分析。结果显示,外观初次检测合格率为60%,性能参数初次检测合格率为96%,初次检测总体合格率为52.5%。经现场或后续故障问题排除后,再次进行检测除颤仪全部合格。由此可知,除颤仪经过长时间使用后性能参数会出现偏差,需要定期进行性能参数的质量检测,才能确保除颤仪保持良好的工作运行状态。 In order to discuss the quality control measures of the performance parameters of defibrillator,40 units,6 brands and 9 models of defibrillator in a grade 3A hospital are used as the monitoring objects,and the test results of the performance parameters are analyzed.The results show that the qualified rate is 60%,96%and 52.5%respectively.After the spot or follow-up troubleshooting,the detection defibrillator is all qualified again.It can be seen that the performance parameters of the defibrillator will deviate after long time use,and the quality inspection of the performance parameters should be carried out regularly to ensure the defibrillator protection with good working running status.
作者 万荣文 管燕 黄仁 黄文娟 WAN Rongwen;GUAN Yan;HUANG Ren;HUANG Wenjuan(Hunan Academy of Traditional Chinese Medicine Affiliated Hospital,Changsha,Hunan,410006,China)
出处 《质量安全与检验检测》 2020年第4期101-102,共2页 QUALITY SAFETY INSPECTION AND TESTING
关键词 除颤仪 性能参数 质量管控 Defibrillator Performance Parameters Quality Control
  • 相关文献

参考文献5

二级参考文献36

  • 1张红蕾.除颤仪技术的讨论[J].医疗设备信息,2006,21(10):35-36. 被引量:19
  • 2JJG760-2003心电监护仪检定规程[S].
  • 3Burch AW, Wong CX, Sullivan T, et al.Trends in the use of implantable cardioverter defibrillators in Australia:a 10-year nationwide study from 2000-2009[J]Jntern MedJ,2013,43(8):888-895.
  • 4Sanghera R,Sanders R,Husby M,et al.Development of the subcutaneous implantable cardioverter-defibrillator for reducing sudden cardiac death[J]Annals of the New York Academy of Sciences, 2014,1329(1):1-17.
  • 5JJF1149-2014,,心脏除颤器校准规范[S].
  • 6Link MS,Atkins DL,Passman RS,et al.Part 6: electrical therapies: automated external defibrillators, defibrillation, cardioversion, and pacing: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care[J].Circulation, 2011,123(6): 235.
  • 7李咏雪,贾建革,卞昕.心脏除颤器和心脏除颤监护仪校准规范[Z].
  • 8JJFl149-2014心脏除颤器校准规范[S].北京:中国计量出版社,2014.
  • 9b HeartStart XL Instructions for Use[M].7th ed.USA: Koninklijke Philips Electronics N.V.,2006.
  • 10HeartStart XL+ Instructions for Use[M].2nd ed.USA:Koninklijke Philips Electronics N.V.,2011.