CT低剂量扫描的探讨 被引量:10

Investigation on low-dose CT scanning
在线阅读 下载PDF
导出
摘要 目的:用西门子Volume Zoom CT机扫描CT模体与人体,探讨降低常规CT扫描剂量的可能性。方法:扫描CT模体,用120kV,mAs在260~180(每次递减20)。共5次。再对50例患者随机分为5组,用120kV,mAs在260~180(每次递减20),进行头颅CT扫描,扫描图像由3名高年资放射诊断医师进行盲审,结果与模体扫描结果进行对比分析。结果:当头颅CT扫描时,mAs由260降至220时,CT剂量指数由62.4降为52.8,降幅16%,两者差异有统计学意义,而图像质量无明显下降,但头颅CT扫描的mAs达到180时,图像噪声大,影响诊断,与模体扫描结果相关。结论:在降低mAs值得情况下,受检者所接受的X线剂量明显减少,在本CT机上,头颅扫描在200mAs以上的图像质量可以满足诊断要求。 Objective:To investigate the possibility of low dose CT scanning for Catphan 500 phantom and patients' heads by Volume Zoom CT scanner (Siemens,Germany). Methods:Catphan 500 phantom was scanned for five times by Volume Zoom CT scanner with same tube voltage 120 KV and different tube currents from 180 mAs to 260 mAs with increment of 20 mAs. 50 patients undergoing head CT scan were enrolled into five groups with same scanning protocols as Catphan 500 phantom (same tube voltage 120 KV and different tube currents from 180 mAs to 260 mAs with increment of 20 mAs). Images were analyzed by three experienced radiologists blindly. Resluts of Catphan 500 phantom and patients heads were compared. Results:Comparison of head CT and Catphan 500 phantom was without statistical difference. For head CT scan,reduction of tube currents from 260 mAs to 200 mAs without image quality imparement,led to statistical decreasing of CTDI from 62.4 to 52.8 with a percentage of 16. With tube current reducing to 180 ms,image noise was too big to make proper diagnosis. Conclusion:X-ray dosage exposure to patients decreases remarkably with tube current reduction. By Volume Zoom CT scanners,image quality of head with tube voltage 120 kV and tube current 200 mAs is sufficient for diagnosis.
出处 《中国医学装备》 2011年第7期36-38,共3页 China Medical Equipment
关键词 CT模体 CT剂量指数 影像质量 低剂量 Catphan 500 phantom Computed tomography dose index Image quality Low dose
  • 相关文献

参考文献6

二级参考文献47

共引文献100

同被引文献125

引证文献10

二级引证文献91

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部