CT机质量控制检测的探讨 被引量:24

The study of CT scanner quality control test
在线阅读 下载PDF
导出
摘要 目的:CT机质量控制检测方法的探讨及应用。方法:用美国模体实验室的Catphan 500模体和瑞典奥利科公司的Solidose 400剂量仪检测CT机扫描架的定位光精度、扫描床运动精度、层厚、CT值线性、视野均匀性、噪声、高对比度分辨率、低对比度分辨率和CT剂量指数(CTDI)。结果:所检测CT机作为整机均合格,但部分指标不合格。结论:通过对噪声、CT线性、对比度分辨率等图像性能参数的检测,可有效保证系统性能良好和维持最优化的图像质量;对CTDI的检测,可及时了解辐射危险水平,进而优化扫描方案,在保证诊断的前提下降低病人受照剂量。 Objective:To study and apply the method of QC test for CT scanners. Methods:We measured the following performance parameters for CT scanners : scan localization light accuracy, table movement accuracy, slice width, CT value linearity and contrast scale, field uniformity, noise, high contrast resolution, low contrast resolution, CT dose index(CTDI) with Catphan 500 phantom made by the phantom laboratory of U. S. A and multi-purpose dosimeter made by Swedishalk Corporation. Results: All CT scanners as appraised are qualified as the whole, but part of mechanic parameters are beyond the regulated limits. Conclusion: To assure the performance of CT system and maintain the optimized image quality according to the image quality parameters test, such as noise, CT value linearity, contrast resolution et al. It is important to understand the radiation danger level and optimize the scanning protocol by checking up the CTDI.
出处 《医学影像学杂志》 2006年第9期978-981,共4页 Journal of Medical Imaging
关键词 体层摄影术 X线计算机 质量控制 检测方法 Tomography, X-ray computed Quality control(QC) Test methods
  • 相关文献

参考文献9

  • 1McCullough EC,Payne JT,Baker HL,et al.Performance evaluation and quality assurance of computed tomography scanners with illustrations from the EMI,ACTA,and delta scanners[J].Radiology,1976,120:173-188.
  • 2White DR,Speller RD,Taylor PM.Evaluating performance characteristics in computerized tomography[J].Br J Radiol,1981,54:221-231.
  • 3AAPM Report No.39,Specification and acceptance testing of computed tomography scanners[J] American Institute of Physics,1993.
  • 4AAPM Report No.74,Quality control in diagnostic radiology,Medical Physics Publishing,2002.
  • 5大型医用设备配置与应用管理暂行办法-卫生部令第43号,1995,7.
  • 6X射线计算机体层摄影装置(CT)应用质量检测与评审规范-全国大型医用设备应用技术评审委员会编,1997.
  • 7余晓锷,林意群,杨金城,陈红文.不同螺距对螺旋CT图像噪声影响的实验研究[J].中国医学物理学杂志,1999,16(4):218-219. 被引量:17
  • 8尉可道.CT剂量的测量及其表达[J].中华放射医学与防护杂志,1997,17(5):347-351. 被引量:43
  • 9尉可道.再论CT剂量的表述[J].中华放射医学与防护杂志,2004,24(2):178-180. 被引量:13

二级参考文献11

  • 1孙良洲.医用X线CT图象质量评价与控制[J].中国医疗器械杂志,1995,19(5):303-307. 被引量:6
  • 2GB18871-2002.电离辐射防护与辐射源安全基本标准.[S].,..
  • 3Wang G,Med Phys,1997年,24卷,1635页
  • 4孙良洲,中国医疗器械杂志,1995年,19卷,303页
  • 5Wang G,Med Phys,1993年,20卷,1635页
  • 6IAEA. International basic safety standards for protection against ionizing radiation and for the safety of radiation sources. Safety series 115, Vienna: IAEA, 1996.
  • 7Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission X-ray computed tomography, Med Phys, 1981,8:488-495.
  • 8EURI. European guidelines on quality critria for computed tomography.EURI, 1999,6262.
  • 9McCrohan JL, Patterson JF, Gagne RM, et al. Average radiation doses in a standard head examination for 250 CT systems. Radiology, 1987,163:263-268.
  • 10ICRP. Managing patient dose in computed tomography, ICRP publication, 2000.87.

共引文献68

同被引文献156

引证文献24

二级引证文献140

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部