胸部肿瘤调强放疗中摆位误差的原因与控制 被引量:10

Set-up Error Analysis and Quality Control of Chest Tumor Cases With IMRT
在线阅读 下载PDF
导出
摘要 目的:测定胸部肿瘤在调强放射治疗中的摆位偏差,分析计划设计中从临床靶区到计划靶区的外扩边界。方法:随机抽取115名胸部肿瘤患者,在治疗时用电子射野影像装置拍摄射野片,将射野片和计划系统中的数字重建射野图像片进行误差比较。结果:在左右、头脚、腹背方向的摆位误差分别是(-0.46±0.24)cm,(-0.37±0.08)cm,(-0.42±0.19)cm,外扩边界分别是0.7 cm,0.45 cm,0.61 cm。结论:对于胸部调强放疗的患者,CTV到PTV的外放边界在左右方向需要2.5 mm,头脚方向和腹背方向需要2 mm。 Objective: To measure set-up error of chest cases in intensity modulated radiation therapy(IMRT).The results can be used to determine the margin from clinical target volume(CTV) to planning target volume(PTV) in treatment planning.Methods: 115 chest tumor cases were randomly chosen,each of them was checked with electronic portal imaging device(EPID).These images were compared with digital reconstructed radiography(DRR) which were generated from treatment planning system.Results: The set-up error in left-right,superior-inferior,anterior-posterior was(-0.46±0.24)cm,(-0.37±0.08)cm,(-0.42±0.19)cm,separately.And the margin in these three directions should be 0.7 cm,0.45cm,0.61cm.Conclusions: For chest cases in radiotherapy,the margin from CTV to PTV should be 2.5 mm in left-right direction and 2mm in both superior-inferior and anterior-posterior directions.
机构地区 浙江省肿瘤医院
出处 《中国医学物理学杂志》 CSCD 2011年第4期2716-2717,2732,共3页 Chinese Journal of Medical Physics
关键词 胸部肿瘤 摆位误差 电子射野影像装置 数字重建图像 chest tumor set-up error EPID DRR
  • 相关文献

参考文献6

  • 1于龙珍,翟振宇,沈君姝.体部肿瘤精确放疗摆位误差分析[J].中国医学物理学杂志,2007,24(3):160-162. 被引量:36
  • 2Commission on Radiation Units and Measurements ( ICRU ). Reports 62. . 1999
  • 3Onishi H,Kuriyama K,Komiyama T,et al.CT evaluation of patient deep inspiration self-breath-holding: how precisely can patients reproduce the tumor position in the absence of respiratory monitoring devices. Medical Physics . 2003
  • 4Stroom JC,de Boer JC,Huizenga H,et al.Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. International Journal of Radiation Oncology, Biology, Physics . 1999
  • 5International Commission on Radiation Units and Measurements.ICRU Report 62: Prescribing, recording, and reporting photon beam therapy (Supplement to ICRU Report 50). ICRU Report 62 . 1999
  • 6Hurkmans CW,Remeijer P,Lebesque JV,et al.Set-up verification using portal imaging; review of current clinical practice. Radiotherapy and Oncology . 2001

二级参考文献8

  • 1Uematsu M,Akira S,Kazuhiko T,et al..Focal,high dose,and fractionated modified stereotactic radiationtherapy for lung carcinoma patients[J].Cancer,1998,82:1062-1070.
  • 2MeertensH,BijholdJ,StrackeeJ.A method of the measurement of field placement errors in digital portal images[J].Phys Med Biol,1990,35:299-323.
  • 3Hurkmans C W,Remeijer P,Lebesque J V,et al.Set-up verification using portal imaging; review of current clinical practice[J].Radiather Oncol,2001,58(2):105-120.
  • 4Hans CJ,John R.Analysis and reduction of 3D systematic and random setup deviations during the simulation and treatment of lung cancer patients with CT based external beam radiotherapy dose planning[J].Int Radiat Oncol Biol Phys,2001,49:857-868.
  • 5Lax I,Blomgren H,Naslund I,et al.S tereotactic radiotherapy of malignacies in the abdomen[J].Acta Oncologia,1994,33:677-703.
  • 6Onishi H,Kuriyama K,Komiyama T,et al.CT evaluation of patient deep inspiration self-breath-holding:how precisely can patients reproduce the tumor position in the absence of respiratory monitoring devices?Med Phys[J].2003,30(6):1183-1187.
  • 7Balter JM,Ten-Haken RK,Lawrence TS,et al.Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing[J].Int Radiat Oncol Biol Phys,1996,36:167-174.
  • 8戴建荣,胡逸民.电子射野影像系统[J].中国医学物理学杂志,1999,16(4):204-208. 被引量:26

共引文献35

同被引文献98

引证文献10

二级引证文献75

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部