摘要
探究了超导3.0TMRI设备和永磁0.4TMRI设备图像均匀性和信噪比检测的对比,分析了ACR体模和SMR170体模检测结果的优劣。对于图像均匀性检测,ACR体模具有更高的灵敏度、真实度和更简便的检测流程。对于信噪比检测,需有效结合MRI设备性能,选择更合适的体模,建立相应的信噪比处置界限。
To explore the comparison of implementing detection for Image Uniformity and Signal-to-Noise Ratio between 3.0T superconducting MRI equipment and 0.4T permanent MRI equipment,and analyze the differences of test results between ACR phantom and Magphan SMR170 phantom.ACR phantom has higher sensitivity,authenticity and more convenient detection process in testing UN.For SNR detection,it is necessary to effectively combine of the equipment performance,select the more appropriate phantom and establish the action limits of SNR.