基于不同体模的MRI图像均匀性和信噪比检测对比研究

Comparative Research of MRI Image Uniformity and Signal-to-noise Ratio Detection Based on Different Phantoms
在线阅读 下载PDF
导出
摘要 探究了超导3.0TMRI设备和永磁0.4TMRI设备图像均匀性和信噪比检测的对比,分析了ACR体模和SMR170体模检测结果的优劣。对于图像均匀性检测,ACR体模具有更高的灵敏度、真实度和更简便的检测流程。对于信噪比检测,需有效结合MRI设备性能,选择更合适的体模,建立相应的信噪比处置界限。 To explore the comparison of implementing detection for Image Uniformity and Signal-to-Noise Ratio between 3.0T superconducting MRI equipment and 0.4T permanent MRI equipment,and analyze the differences of test results between ACR phantom and Magphan SMR170 phantom.ACR phantom has higher sensitivity,authenticity and more convenient detection process in testing UN.For SNR detection,it is necessary to effectively combine of the equipment performance,select the more appropriate phantom and establish the action limits of SNR.
作者 郁佳琦 陆翊铭 金文燕 徐盛洋 陈珊珊 YU Jiaqi;LU Yiming;JIN Wenyan;XU Shengyang;CHEN Shanshan
出处 《计量与测试技术》 2021年第12期16-19,共4页 Metrology & Measurement Technique
基金 教育教学改革专项——创新创业训练项目(国家级项目)(项目编号:A3-0200-21-309010-354)。
关键词 MRI 质量控制 图像均匀性 信噪比 MRI quality control image uniformity signal-to-noise ratio