期刊文献+

电化学工作站校准方法 被引量:2

Research on the calibration method of electrochemical workstations
在线阅读 下载PDF
导出
摘要 针对我国目前缺少电化学工作站校准方法的情况,文章通过阐述电化学工作站的工作原理,从仪器的物理性能入手,提出了该仪器的校准项目和技术指标:电位示值误差(±0.02 V)、电流示值误差(±5×10^(-9) A)、峰电位重复性(3%);重点探讨了仪器的电位示值误差、电流示值误差和峰电位重复性指标的校准方法,形成了一套系统的电化学工作站校准方法,并对电化学工作站进行了实际校准。结果表明该方法科学合理、简便易行、可操作性强,各项指标结果符合仪器的校准项目指标,可用于评价仪器计量性能,为仪器日常的校准工作以及国家计量校准规范的制订奠定了基础。 In view of the lack of calibration method for electrochemical workstations in China,the paper described the working principle of the instrument and has proposed the calibration items and technology indicators,including of indication error of potential(±0.02 V),indication error of current(±5×10^(-9) A)and repeatability of peak potential(3%).The calibration method for electrochemical workstations was mainly discussed and carried out on two electrochemical workstations.The results were all in line with the technology indicators of the instrument.Analysis of the cases showed that the method was scientific,reasonable,simple and easy to operate.The paper constructed the experimental foundation for the calibration activity and standards for the electrochemical workstations.
作者 杨雪 王乐乐 丁敏 闻艳丽 徐勤 李兰英 Yang Xue;Wang Lele;Ding Min;Wen Yanli;Xu Qin;Li Lanying(Shanghai Institute of Measurement and Testing Technology)
出处 《上海计量测试》 2021年第1期2-5,9,共5页 Shanghai Measurement and Testing
基金 国家质量基础的共性技术研究与应用重点专项(2018YFF0212803) 国家自然科学基金面上基金(21775104)。
关键词 电化学工作站 计量校准 电位示值误差 电流示值误差 峰电位重复性 electrochemical workstations calibration indication error of potential indication error of current repeatability of peak potential
  • 相关文献

参考文献5

二级参考文献19

共引文献31

同被引文献11

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部