双源CT低管电压Flash扫描在孕妇新型冠状病毒肺炎检查中的应用价值 被引量:2

Application value of dual-source CT low-kV Flash scan in screening pregnant women with COVID-19
在线阅读 下载PDF
导出
摘要 目的:探讨双源CT低管电压Flash扫描在孕妇新型冠状病毒肺炎(简称新冠肺炎)筛查中的应用价值。方法:回顾性分析12例孕妇和50例普通患者的胸部CT资料,分为A组、B组。A组采用Siemens双源CT Flash模式对肺部进行扫描,B组采用常规肺部扫描模式,2组管电压80 kV,管电流使用自动毫安秒技术。采用双盲法对2组图像质量进行评分。比较2组CT检查时间、CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP)及有效剂量(ED)。结果:2组图像质量评分差异有统计学意义(χ~2=11.49,P=0.003)。A组扫描时间、CTDIvol、DLP及ED与B组比较分别减少约69.71%、39.56%、41.53%、42.76%,差异均有统计学意义(均P<0.001)。结论:双源CT Flash扫描能够清晰显示患者肺部感染情况,且辐射剂量低,值得在孕妇的新冠肺炎筛查中广泛应用。 Objective:To explore the value of dual-source CT low-kV Flash scanning mode in screening pregnancies with COVID-19.Methods:The chest CT image data of 12 pregnant women and 50 ordinary patients,who were divided into A and B group,in our hospital after January 2,2020 were retrospectively analyzed.Group A used Siemens dual-source CT Flash mode to scan the lungs,Group B used conventional lung scanning mode,and the tube voltage was 80 kV,tube current using automatic milliammeter technology.Two groups of image quality were scored by double-blind method.The scan time of CT examination,volume dose index(CTDIvol),dose length product(DLP),and effective dose(ED) in both groups were recorded in each patient.Results:The image quality of the two groups had a difference (χ~2=11.49,P=0.003).The scan time,CTDIvol,DLP and ED of Group A decreased by about 69.71%,39.56%,41.53%,42.76%,respectively.Conclusion:Double source CT Flash scan can clearly show the pulmonary infection and has lower radiation dose,which can be considered as a worthy application in screening pregnant women with COVID-19.
作者 赵明 刘海峰 张东友 尹龙 ZHAO Ming;LIU Haifeng;ZHANG Dongyou;YIN Long(Department of Radiology,Wuhan No.1 Hospital,Wuhan 430030,China)
出处 《中国中西医结合影像学杂志》 2020年第4期328-330,共3页 Chinese Imaging Journal of Integrated Traditional and Western Medicine
关键词 放射摄影术 双能扫描投影 孕妇 辐射剂量 Radiography dual-energy scanned Pregnant women Radiation dosage
  • 相关文献

参考文献9

二级参考文献59

  • 1郑颖彦,马昕,王慧英,王东东,耿道颖,尹波.新型冠状病毒肺炎的薄层高分辨率计算机断层扫描征象[J].上海医学,2020(5):261-265. 被引量:58
  • 2龚晓明,李航,宋璐,李莉,张笑春,余开湖,鲁植艳,李宏军.新型冠状病毒肺炎(COVID-19)CT表现初步探讨[J].放射学实践,2020,35(3):261-265. 被引量:62
  • 3刘晓红,何仲.德尔菲法在护理研究中的应用[J].中华护理教育,2005,2(1):6-8. 被引量:70
  • 4刘绪明,蒋贤高,邱乾德,姜英,孙厚长.重症及危重症甲型H1N1流感合并肺炎的CT表现及动态变化[J].医学影像学杂志,2010,20(12):1884-1887. 被引量:5
  • 5NTaeelli,MRemy-Jardin,TFlohr,等.高时间分辨率及长螺距模式双源cT胸部血管成像:140例病人影像质量的评价[J].国际医学放射学杂志,2010,4(7):397-398.
  • 6Willis CE, Slovis TL.The ALARA concept in pediatric CR and DR : dose reduction in pediatric radiographic exams-a white paper con- ference [ J :.AJR Am J Roentgenol, 2005,18 (2) : 373-374.
  • 7Kim SM, Haider MA, Milosevic M, et al.A method for patient dose reduction in dynamic contrast enhanced CT study [J]. Med Phys, 201 1,38(9) :5094-5103.
  • 8Heliou R, Normandeau L, Beaudoin G. Towards dose reduction in CT : patient radiation dose assessment for CT examinations at univer- sity health center in Canada and comparison with national diagnos- tic reference levels [J].Radiat Prot Dosimetry, 2012, 48 (2) : 202-210.
  • 9Behrendt FF, Schmidt B, Plumhans C, et al. Image fusion in dual en- ergy computed tomography : effect on contrast enhancement, signl-to-noise ratio and image quality in computed tomography angiography [J:. Invest Radio1,2009,44( 1 ) : 1-6.
  • 10Menzel H, Schibilla H, Teunen D, et al. European guidelines on quality criteria for computed tomography : C ]. Luxembourg: Europe- an Commission,2000,12(2) : 160-162.

共引文献1125

同被引文献23

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部