新型眼生物测量仪Tomey OA-2000和IOLMaster获得的白内障患者眼生物参数的比较 被引量:11

Comparison of ocular variables obtained from Tomey OA-2000 and IOLMaster
在线阅读 下载PDF
导出
摘要 目的比较新型眼生物测量仪Tomey OA-2000与IOLMaster获得的白内障患者眼生物参数的差异性和一致性。方法前瞻性研究。纳入白内障术前检查者74例(74眼),分别应用Tomey OA-2000和IOL-Master进行测量,获得眼轴长度(axial length,AL)、角膜曲率(keratometry,K)、前房深度(anterior chamber depth,ACD)和角膜横径(white to white,WTW)等参数。配对t检验分析两种设备获得的眼生物参数之间的差异。Bland-Altman散点图分析两种设备获得的各生物参数之间的一致性。结果Tomey OA-2000获得的AL、平坦K(flat K,Kf)值、陡峭K(steep K,Ks)值、平均K(mean K,Km)值、中央角膜厚度(central corneal thickness,CCT)、ACD、晶状体厚度(lens thickness,LT)、瞳孔直径(pupil diameter,PD)和WTW值分别为(24.26±2.71)mm、(44.09±1.76)D、(45.01±1.73)D、(44.55±1.72)D、(518.41±34.18)μm、(3.07±0.48)mm、(4.48±0.61)mm、(4.07±0.95)mm和(11.50±0.52)mm。IOLMaster获得的AL、Kf、Ks、Km、ACD和WTW值分别为(24.33±2.69)mm、(44.21±1.74)mm、(45.15±1.76)D、(44.69±1.74)D、(3.04±0.46)mm和(11.65±0.41)mm。Tomey OA-2000和IOLMaster获得的AL、Kf、Ks、Km、ACD和WTW差值分别为(-0.07±0.10)mm(t=-5.99,P<0.01)、(-0.12±0.14)mm(t=1.99,P=0.05)和(-0.14±0.32)mm(t=-3.73,P<0.01)。两种设备获得的AL、Kf、Ks、Km、ACD的95%一致性界限(Lo A)其上下限最大绝对值分别为0.27 mm、0.50 D、0.53 D、0.38 D、0.30 mm,一致性较好;而WTW的95%Lo A为0.78 mm,一致性较差。结论Tomey OA-2000和IOLMaster获得的AL、K、ACD差异小,一致性好。 Objective To compare the ocular parameters measured by 2 biome- try devices Tomey OA-2000 and IOLMaster in cataract eyes. Methods A total of 74 cataract patients (74 eyes) were included in this prospective study. Ocular parameters, including axial length ( AL), keratometry ( K), anterior chamber depth ( ACD ) and white to white (WTW), were obtained from Tomey OA-2000 and IOLMaster, respective- ly. Paired t-test was applied to analyze the differences of ocular parameters from Tomey OA-2000 and IOLMaster,while Bland-Altman plot was used to assess the agreement of these ocular parameters. Results The AL, flat K, steep K, mean K, central corneal thickness (CCT), ACD, lens thickness, pupil diameter and WTW values measured by Tomey OA-2000 were (24.26±2.71 ) mm, (44.09 ± 1.75 ) D, (45.01 ± 1.73 ) D, ( 44.55 ± 1.72)D, (518.41 ±34.18)μm, (3.07±+0.48) mm, (4.48± 0.61) mm, (4.07 ± 0.95) mm and ( 11.50 ±0.52 ) mm respectively. Values of AL, flat K, steep K, mean K, ACD and WTW from IOLMaster were ( 24.33±2.59 ) mm, ( 44.21 ± 1.74 ) mm, ( 45.15 ± 1.75 ) D, (44.69 ± 1.74) D, (3.04 ± 0.45) mm and ( 11.55 ± 0.41 ) nun respectively. The difference values of AL, flat K, steep K, mean K, ACD and WTW between Tomey OA-2000 and IOL- Master were ( -0.07 ±0. 10) mm (t = -5.99,P 〈0. 01), ( -0. 12 ± 0.21) D (t = -4.14,P〈0.01),(-0.14 ± 0.20)D (t= -5.88,P〈0.01),(-0.12±0.13) D (t= -7.44,P〈0.01) ,(0.03 ±0.14) mm (t = 1.99,P=0.05) and ( -0.14 ±0.32) nun (t= - 3.73 ,P 〈 0.01 ) respectively. The 95% limits of agreement (LoA) of the AL, flat K, steep K,mean K, ACD obtained from the two devices was 0.27 mm,0.50 D,0.53 D, 0. 38 D,0. 30 nun, respectively, and the agreement was good; 95% LoA of WTW was 0.78 nun, and the consistency was poor. Conelusion The differences of AL,K values and ACD between Tomey OA-2000 and IOLMaster were small, suggesting the agreement is unified.
出处 《眼科新进展》 CAS 北大核心 2017年第9期845-848,共4页 Recent Advances in Ophthalmology
基金 上海市卫生与计划生育委员会重点项目基金资助(编号:201440029)~~
关键词 Tomey 0A-2000 IOLMASTER 眼轴长度 前房深度 角膜曲率 Tomey OA-2000 IOLMaster axial length anterior chamber depth ker-atometry
  • 相关文献

参考文献7

二级参考文献110

共引文献77

同被引文献77

引证文献11

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部