期刊文献+

四肢血压同步测量仪研制原理与医学实践 被引量:8

DEVELOPMENT OF THE SYNCHRONOUS SPHYGMOMANOMETER FOR FOUR LIMBS—TECHNICAL BASICS AND MEDICAL PRACTICE
在线阅读 下载PDF
导出
摘要 本文从医学计量工作的实践出发 ,谈了血压、血压计以及参与了“电子血压计准吗 ?”的讨论。经过大量比对测量 ,认为 :袖带式、合格的电子血压计测量的血压基本上是准确的。这就是本项课题研究的技术基础采用四个技术参数相近的、全自动电子血压计为基础组件 ,研制出四肢血压同步测量仪。经过大量实测发现 :人在水平仰卧时 ,同步测得的四肢血压值之间相差得越大 ,与血液粘稠度高、高脂血症、血管内粥状血栓形成、血管硬化等 ,越密切相关 ,为今后临床检测与有关疾病的相关性研究打下了良好的基础。 The accuracy of the qualified cuff-type autoelectronic sphygmomanometer, which was the technical foundation of this study, had been confirmed by many comparative measurements. The prototype of the 'Synchronous Sphygmomanometer for Four Limbs' was developed based on four auto-electronic sphygmomanometers of similar parameters. The results from our experiments demonstrated that the differences of blood pressure among the four limbs under the supine position were highly correlated with the high density of blood, hyperlipemia, congee-like thrombosis and the hardening of blood vessels. These findings may be useful in the studies of the association between clinical blood pressure measurements and related diseases.
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 2002年第2期182-186,共5页 Chinese Journal of Biomedical Engineering
关键词 四肢血压 同步测量 血压计 Blood pressure of four limbs synchronous measurement sphygmomanometer
  • 相关文献

参考文献3

  • 1美国标准:Electronic or Automated Sphygmomanometers[S].ANSI/AAMI SP10-1992.
  • 2英国高血压协会(BHS).关于以听诊法为参照标准进行自动和半自动血压测量仪评估的协议[S].1996.
  • 3日本标准:非观血式电子血压计[S].JJS T1115-1987.

同被引文献77

  • 1曹柏荣,瞿丹晨,冯运达,顾宏奎.人体柯氏音测量方法的研究[J].仪表技术,2004(6):36-37. 被引量:7
  • 2邹峰.脉搏波信号处理的一种简单可靠的实时算法[J].生物医学工程学杂志,1994,11(2):182-183. 被引量:2
  • 3陈龙聪,胡国虎.计数型模拟量及人体血管壁动态信息多道遥测电脑系统的研制[J].生物医学工程学杂志,2006,23(1):200-204. 被引量:4
  • 4张瑾,魏菁,席惠芳,何清波,朱鼎良.HEM-6000手腕式血压计的可靠性[J].中华高血压杂志,2007,15(5):407-410. 被引量:9
  • 5扬子彬.当代医学新理论与新技术丛书:基础医学卷[J].生物医学工程学,2000,.
  • 6陈芝村.动脉血压调节机理及测量[M].北京人民军医出版社,1987..
  • 7Association for the Advancement of Medical Instrumentation, American national standard. Electronic or automated sphygmomanometers [S]. ANSI/AAMI SP 10-1992, Arlington, VA: AAMI, 1993: 1-40.
  • 8Hannu Sorvoja, Risto Myllyla. Noninvasive blood pressure measurement methods [J]. Molecular cmd Quantum Acoustics, 2006, 27: 239-264.
  • 9Geddes LA. The direct and indirect measurement of blood pressure[M]. Chicago: Year Book Medical Publishers, 1970: 82-130.