Z-G旁流式EtCO_2监测吸氧装置的准确性及其临床应用

Accuracy and clinical effect of Z-G sidestream non-intubation oxygen device for monitoring EtCO_2
在线阅读 下载PDF
导出
摘要 目的评价Z-G旁流式EtCO_2监测吸氧装置EtCO_2监测准确性及其在非气管插管全身麻醉中的应用效果。方法乳房肿块切除女性患者100例,予非气管插管全身麻醉,术中使用旁流式非气管插管EtCO_2监测吸氧装置。予咪达唑仑、芬太尼、异丙酚行静脉麻醉保留自主呼吸,记录给药前(T_0)、给药后EtCO_2波形稳定后5 min(T_1)、EtCO_2波形稳定后10min(T_2)的MAP、HR、SpO_2、RR、P_(ET)CO_2;分别于各时间点抽取动脉血行血气分析。结果麻醉前P_(ET)CO_2为(33.25±1.34)mmHg,p_aCO_2为(36.20±1.15)mmHg。术中P_(ET)CO_2、P_aCO_2、pH在T_1、T_2显示轻度的呼吸性酸中毒的表现。P_(ET)CO_2与P_aCO_2相关系数r=0.627,呈正的直线相关,Pa-etCO_2平均为3.77 mmHg。所有患者的点在95%一致性界限以内,一致性界限以内Pa-etCO_2最大值为6.2 mmHg,P_(ET)O_2和P_aCO_2的平均值为38.5 mmHg。结论 Z-G旁流式EtCO_2监测吸氧装置监测EtCO_2使用方便,P_(ET)CO_2与P_aCO_2相关性好,可以及早、迅速、准确的发现呼吸异常,值得推荐使用。 Objective To evaluate the accuracy and clinical effect of Z-G sidestream non- endotracheal intubation oxygen device in monitoring EtCO2. Methods One hundred female patients underwent breast lump resection. The non-endotracheal intubation anesthesia and the Z-G sidestream non-endotracheal intubation oxygen device were applied in operation. Midazolam, fentanyl and propofol was given for intravenous anesthesia with spontaneous respiration and MAP, HR, SpO2, RR, PETCO2 were recorded immediately before administration (T0) ; 5 min (T1 ) and 10 min (T2 ) after the waveform of EtCO: was stable. At each time point, arterial blood was collected for blood gas analysis. Results PETCO2 was (33.25 ±1.34) mmHg, paCO2 was (36.20 ±1.15 ) mmHg before anesthesia. Intra-operative PETCO2. paCO2, pH at T1and T2 showed slight respiratory acidosis. The correlation coefficient of PETCO2 and PaCO2 was 0. 627 with a positive linear correlation ; the average value of Pa- etCO2 was 3.77 mmHg. All points were in 95% consistency within boundaries, the maximal Pa-etCO2 was 6. 2 mmHg within boundaries, the average value of PET CO2 and pa CO2 was 38. 5 mmHg. Conclusion Both the effect of monitoring and offering oxygen is exact when using the Z-G sidestream non-intubation device. The abnormal breathing can be found early, rapidly and accurately when using this device. PETCO2 is positively correlated with paCO2. EtCO2 and p, CO2 has good consistency and high accuracy, so it is worth promoting.
出处 《同济大学学报(医学版)》 CAS 2013年第4期68-71,共4页 Journal of Tongji University(Medical Science)
关键词 旁流式 静脉全麻 非气管插管 P_ETCO_2 p_aCO_2 sidestream intravenous anesthesia non-intubation PETCO2 paCO2
  • 相关文献

参考文献8

  • 1许小婷,陈彦青.监护性麻醉下经鼻咽通气管监测呼气末二氧化碳分压的准确性[D].福州:福建医科大学,2010.
  • 2张春艳,王儒蓉.清醒患者经微流口鼻罩EtCO2监测的准确性[D].成都:四川大学华西医院,2012.
  • 3Zhang C,Wang M, Wang R,et al. Accuracy of end-tidal C02 measurement through the nose and pharynx innonintubated patients during digital subtraction cerebralangiography [ J ]. J Neurosurg Anesthesiol, 2013,25(2): 191-196.
  • 4Miller RD.米勒麻醉学[M ].邓小明,曾因明,译.北京:北京大学医学出版社,2011: 1437.
  • 5Lee SW, Hong YS, Han C, et al. Concordance ofend-tidal carbon dioxide and arterial carbon dioxide insevere traumatic brain injury [ J]. J Trauma,2009,67(3): 526-530.
  • 6李燕,代志刚,张示杰.持续PetCO_2与PaCO_2相关性研究进展[J].农垦医学,2008,30(6):500-503. 被引量:1
  • 7李燕,代志刚,张示杰.持续呼气末二氧化碳分压与动脉二氧化碳分压相互的关系[J].石河子大学学报(自然科学版),2010,28(1):76-78. 被引量:13
  • 8殷曰昊,朱敏敏,谭志明,陈家伟,徐亚军,顾悦超.熵指数监测短小手术保留自主呼吸非气管插管全麻的气道管理[J].中国临床医学,2011,18(3):383-385. 被引量:1

二级参考文献25

共引文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部