黄疸因素对免疫透射比浊法检测血清白蛋白的影响 被引量:1

Influence of jaundice on immune transmission turbidimetry in detection of serum albumin
在线阅读 下载PDF
导出
摘要 目的探讨免疫透射比浊法中黄疸对检测血清白蛋白结果的影响。方法将双份已知白蛋白浓度(低、高值)的血清中分别加入胆红素纯品,制成含不同胆红素浓度梯度的黄疸血清,再用免疫透射比浊法测定其白蛋白含量,比较分析黄疸因素对血清白蛋白检测的影响。结果普通黄疸(胆红素浓度低于400umol/L)对白蛋白检测结果无显著性差异(P>0.05)。结论选用免疫透射比浊法检测血清白蛋白对黄疸的抗干扰能力较强。 Objective To investigate the influence of jaundice on the detection ot serum albumin Dy immune transmission turbidimetry. Methods Pure bilirubin was added to two samples of serum with known albumin concentrations (low and high values) in order to make the jaundice serum with different bilirubin concentration gradients. Then the immune transmission turbidimetry was used to determine the albumin content and the influence of jaundice on serum albumin detection was analyzed. Results Common jaundice (with a bilirubin concentration of less than 400 umol/L) had no significant influence on the albumin detection results (P 〉 0.05). Conclusion Immune transmission turbidimetry has strong anti-jamming capability against jaundice in the detection of serum albumin.
出处 《中国医药科学》 2013年第12期122-123,共2页 China Medicine And Pharmacy
关键词 黄疸 免疫透射比浊法 白蛋白 抗干扰 Jaundice Immune transmission turbidimetry Albumin Anti-jamming
  • 相关文献

参考文献6

二级参考文献21

  • 1张汉园.透射免疫比浊法和速率散射免疫比浊法测定C反应蛋白的比较[J].江苏大学学报(医学版),2004,14(4):348-348. 被引量:18
  • 2杨春平.溧阳地区血清总蛋白和白蛋白及A/G比值参考值调查[J].南京军医学院学报,2003,25(4):292-293. 被引量:9
  • 3文庆成,马志荣,师本章.干化学分析在医学检验中的应用[J].辽宁医学杂志,1995,9(4):171-173. 被引量:35
  • 4伍绍国,李志海,鲍蓓,杜利军,何妙贤.广州成人血清总蛋白、白蛋白及A/G比值参考值调查[J].现代临床医学生物工程学杂志,2006,12(2):135-137. 被引量:19
  • 5Blirup-Jensen S. Protein standardization lit :Method optimization basic principles for quantitative deter- mination of human serum proteins on automated in- struments based on turbidimetry or nephelometry [J]. Clin Chem Lab Med,2001,39(11) : 1098-1109.
  • 6Ledue TB,Collins MF,Ritchie RF. Development of immunoturbidimetric assays for fourteen human serum proteins on the Hitachi 912^TM[J]. Clin Chem Lab Med,2002,40(5) :520-528.
  • 7National Committee for Clinical l~boratory Standards. Method comparison and bias estimation using patient samples. Approved guideline, NCCLS, EP9-A, 1995.
  • 8National Committee for Clinical Laboratory Standards. Method comparison and bias estimation using patient samples. Approved guideline, NCCLS, EP9-A, 1995
  • 9Avram MM,Goldwasser P,Erroa M,et al.Predictors of survival in continuous ambulatory peritoneal dialysis patients:the importance of prealbumin and other nutritional and metabolic markers[J].Am J Kidney Dis,1994,23:91-98.
  • 10Agbedana EO,Oshota JO,Taylor GO,et al.Problems associated with plasma albumin estimation in nephrotic syndrome using the bromocresol green method[J].Afr J Med Med Sci,1993,22:45-48.

共引文献264

同被引文献7

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部