ArcCHECK半导体探头特性及在容积调强弧形治疗剂量验证应用研究 被引量:28

The characteristics and clinical application of the ArcCHECK diode array for volumetric-modulated arc therapy verification
原文传递
导出
摘要 目的研究ArcCHECK半导体探头特性及在容积调强弧形治疗(VMAT)剂量验证应用。方法用PTWRW3型固体水模体对ArcCHECK探头测量固有敏感性、稳定性、剂量响应、剂量率响应、每脉冲剂量响应、射野大小依赖性、角度依赖性,并与PTW31010型0.125cm。指形电离室测量值或VMAT计划系统计算值比较。随机选取211例已验证过的VMAT计划,分析计划与测量剂量分布的y通过率差异,两两比较采用配对t检验。结果除角度依赖性外ArcCHECK探头其余特性均符合临床验证要求,当射束从探头底部入射时探头响应最小(180。时约为一3.9%),射束从两侧入射时响应最大(255。时约为7.7%)。113例鼻咽癌、48例宫颈癌和50例直肠癌VMAT计划3mm3%的y平均通过率分别为93.5%、95.7%和97.5%,两两比较t=一12.69~一4.88,P均〈0.01。结论ArcCHECK半导体探头进行VMAT剂量验证前需精心校正,计划复杂度是影响VMAT计划验证通过率主要因素。 Objective To evaluate the characteristics and clinical suitability of the ArcCHECK diode array for volumetric-modulated arc therapy (VMAT) verification. Methods The intrinsic sensitivity, short and long term reproducibility, dose and dose rate dependence, dose per pulsed dependence, field sizes dependence and directional response of the diodes were measured. The results of the diodes were compared with the measurement results of an ionization chamber and calculated results of treatment planning system. Gamma index was used to analyze the dose difference between the calculation and measurement for random selected 211 verified VMAT plans. Results The ArcCHECK performed well for all tests except directional dependence, which varies from a minimum of - 3.9% ( seen only when the beam was incident on the diode at 180° to a maximum of 7. 7% (approximately at 255°). Average gamma analysis passing rates with 3 mm/3% for 113 nasopharyngeal cancer, 48 cervical cancer and 50 rectal cancer patients VMAT plans were 93.5% ,95.7% and 97. 5%, respectively, statistical significance has been found between either two of the groups ( t = - 12. 69-- -4. 88, all P 〈 0. 01 ). Conclusions With proposed calibration method, the ArcCHECK diode array is very suitable for VMAT pretreatment verification, the complexity of VMAT plan is the main factor affecting the passing rate.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2013年第3期253-257,共5页 Chinese Journal of Radiation Oncology
基金 国家自然科学基金项目(81101697)
关键词 ArcCHECK半导体探头 容积调强弧形治疗 剂量验证 ArcCHECK semiconductor probe Volumetric-modulated arc therapy Dose verification
  • 相关文献

参考文献23

  • 1Otto K. Volumetric modulated arc therapy:IMRT in a single gantry arc. Med Phys,2008,35:310-317.
  • 2Bortfeldand T, Webb S. Single-Arc IMRT? Phys Med Biol,2009,54:19-20.
  • 3Wang C, Luan S, Tang G, et al. Arc-modulated radiation therapy (AMRT):a single-arc form of intensity-modulated arc therapy. Phys Med Biol,2008,53:6291-6303.
  • 4Bedford JL, Warrington AP. Commissioning of volumetric modulated ARC therapy (VMAT). Int J Radiat Oncol Biol Phys,2009,73:537-545.
  • 5Ling CC, Zhang P, Archambault Y, et al. Commissioning and quality assurance of RAPIDARC radiotherapy delivery system. Int J Radiat Oncol Biol Phys,2008,72:575-581.
  • 6Low DA, Moran JM, Dempsey JF, et al. Dosimetry tools and techniques for IMRT. Med Phys,2011,38:1313-1338.
  • 7Rikner G, Grusell E. General specifications for silicon semiconductors for use in radiation dosimetry. Phys Med Biol,1987,32:1109-1117.
  • 8Robar JL, Clark BG. The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry. Med Phys,1999,26:2144-2150.
  • 9Yeo IJ, Beiki-Ardakani A, Cho YB, et al. EDR2 film dosimetry for IMRT verification using low-energy photon filters. Med Phys,2004,31:1960-1963.
  • 10Jursinic PA, Nelms BE.A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. Med Phys,2003,30:870-879.

同被引文献139

  • 1张矛,金海国,卜明伟,孙宝胜,孙术全,苏清秀,李玉平.脑胶质瘤术后VMAT与IMRT放疗技术间比较[J].中国医学物理学杂志,2011,28(6):2959-2963. 被引量:30
  • 2戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 3王新卓.良性脑膜瘤术后放射治疗的研究[J].国外医学(临床放射学分册),2005,28(2). 被引量:3
  • 4黄晓延,黄劭敏,张黎,邓小武,钟宁山,陈立新.三维治疗计划系统的剂量学验证[J].中华放射肿瘤学杂志,2006,15(6):496-500. 被引量:26
  • 5马林,王连元.周桂霞.肿瘤断层放射治疗[M].成都:四川科学技术出版社,2010:55.
  • 6Lang S, Reggiori G, Puxeu-Vaquee J, et al. Pretreatment quality assurance of flattening filter free beams on 224 patients for intensity modulated plans: a multicentric study [J]. Med Phys, 2012, 39 (3) :1351-1356. DOI:10.1 118/1. 3685461.
  • 7Li G,Zhang Y,Jiang X. et al. Evaluation of the ArcCHECK QA system for IMRT and VMAT verification [J]. Phys Med,2013,29: 295-303. DOI: 10. 1016/j. ejmp. 2012.04. 005.
  • 8Ezzell GA, Burmeiste JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM task group 119 [ J ]. Med Phys, 2009,36 ( 11 ) :5371 - 5372.
  • 9MeGarry CK, O' Connell BF, Grattan MW, et al. Octavius 4D characterization for flattened and flattening filter free rotational de- liveries [J]. Med Phys,2013,40(9):l-ll. DOI:I0. 1 118/1. 4817482.
  • 10Molinier J, Vieillevigne L, Brun T, et al. Comparison of three meas- urement systems for RapidArc treatment verification [J]. Phys Med,2012,28(Suppl 1) :S13.DOI: 10.1016/j.ejmp.2012.08.055.

引证文献28

二级引证文献120

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部