摘要
QUASAR Penta-guide模体粘贴定位标记,CT扫描图像传输工作站进行靶区勾画,用Tomo Plan系统设计计划。模体进行MVCT扫描,图像配准后治疗床自动位置调整,绿激光灯在左右和头脚方向与中心长十字线重合,误差小于2mm;垂直方向存在床面下沉,与中心长十字线有2mm的偏差,校正后能消除影响。头部或盆腔肿瘤患者,MVCT扫描调整治疗床,再次扫描配准分析残存误差。保证影像配准后治疗床位置校正准确。
The QUASAR Penta-guide Phantom with fiducial markers was scanned, and the CT images were transferred to Pinnacle workstation. Skin and target volumes were contoured and transferred to TomoPlan treatment planning system. The phantom was scanned with Megavoltage CT (MVCT). MVCT images were matched to the planning CT. Automatic adjustment of treatment couch was completed. It was found that the green laser coincided with the etched center crosshairs in lateral and longitudinal directions with an error less than 2 mm. However 2 mm vertical tabletop lag was found, but could be eventually corrected. Verifications for specific patients with head and pelvic tumors were also completed, the residual setup error were analyzed. The automatic movement of tabletop after image match is satisfactory.
出处
《中国医疗器械杂志》
CAS
2013年第2期143-145,共3页
Chinese Journal of Medical Instrumentation