期刊文献+

压握过程对球囊扩张支架性能的影响 被引量:6

Influence of Crimping on the Performances of Balloon Expandable Stents
在线阅读 下载PDF
导出
摘要 压握过程中支架发生塑性变形及加工硬化现象将对支架的后续变形行为产生影响.研究利用有限元方法模拟了支架的完整变形过程,评价了扩张后支架的各项性能,并与未考虑压握过程的模拟结果进行了比较.结果表明,压握过程不仅会导致扩张后支架的形状发生变化,同时会使回弹率和最大应力增大,支架最大可扩张直径变小. Plastic deformation and work hardening may occur in crimping, which can result in changes of material properties and affect the performance of stents. A complete deformation procedure of a typical stent was simulated using finite element method. Its mechanical performances were evaluated and com- pared with that using a simplified procedure (omitting the process of stent crimping). It was found that the crimping has a significant influence on the stent shape, stress distribution, recoil rate and the maximum expandable diameter of stent.
机构地区 上海交通大学
出处 《上海交通大学学报》 EI CAS CSCD 北大核心 2012年第4期646-650,共5页 Journal of Shanghai Jiaotong University
基金 国家自然科学基金资助项目(30970704)
关键词 压握 加工硬化 球囊扩张支架 crimping work hardening balloon expandable stent
  • 相关文献

参考文献6

  • 1李田昌.冠心病介入治疗现状[J].中国全科医学,2007,10(16):1317-1320. 被引量:20
  • 2Kiousis D E, Wulff A R, Holzapfel G A. Experimen- tal studies and numerical analysis of the inflation and interaction of vascular balloon catheter-stent systems [J]. Annals of Biomedical Engineering, 2009, 37 : 315-330.
  • 3Migliavacca F, Petrini L, Montanari V, et al. A pre- dictive study of the mechanical behavior of coronary stents by computer modeling [J].Medical Engineering & Physics, 2005, 27(1):13-18.
  • 4倪中华,王跃轩,程洁.球囊扩张式冠脉支架扩张变形机理数值模拟方法[J].机械工程学报,2008,44(1):102-108. 被引量:19
  • 5Zahedmanesh H, Kelly D J, Lally C. Simulation of a balloon expandable stent in a realistic coronary artery - Determination of the optimum modeling strategy [J]. Journal of Biomechanics, 2010, 43:2126-2132.
  • 6Chua S N D, Macdonald B J, Hashmi M S J. Effects of varying slotted tube (stent) geometry on its expan- sion behavior using finite element method [J]. Materials Processing Technology, 2004, 155-156:1764-1771.

二级参考文献9

  • 1倪中华,易红,顾兴中.载药纳米颗粒与血管支架自装配机理和方法[J].机械工程学报,2005,41(8):190-194. 被引量:4
  • 2LALLY C,DOLAN F,PRENDERGAST P J.Cardiovascular stent design and vessel stresses:a finite element analysis[J].Journal of Biomechanics,2005,38(8):1 574-1 581.
  • 3WALKE W,PASZENDA Z,FILIPIAK J.Experimental and numerical biomechanical analysis of vascular stent[J].Journal of Materials Processing Technology,2005,164-165:1 263-1 268.
  • 4CHUA S N D,MACDONALD B J,HASHMI M S J.Finite element simulation of slotted tube (stent) with the presence of plague and artery by balloon expansion[J].Materials Processing Technology,2004,155-156:1 772-1 779.
  • 5CHUA S N D,MACDONALD B J,HASHMI M S J.Effects of varying slotted tube (stent) geometry on its expansion behavior using finite element method[J].Materials Processing Technology,2004,155-156:1 764-1 771.
  • 6MIGLIAVACCA F,PETRINI L,MONTANARI V,et al.A predictive study of the mechanical behavior of coronary stents by computer modeling[J].Medical Engineering & Physics,2005,27(1):13-18.
  • 7MIGLIAVACCA F,PETRINI L,COLOMBO M.Mechanical behavior of coronary stents investigated through the finite element method[J].Journal of Biomechanics,2002,35(6):803-811.
  • 8PETRINI L,MIGLIAVACCA F,AURICCHIO F,et al.Numerical investigation of the intravascular coronary stent flexibility[J].Journal of Biomechanics,2004,37(4):495-501.
  • 9AURICCHIO F,LORETO M D,SACCO E.Finite element analysis of a stenotic artery revascularization through a stent insertion[J].Computer Methods in Biomechanics and Biomedical Engineering,2001,4(5):249-263.

共引文献36

同被引文献43

引证文献6

二级引证文献8