期刊文献+

螺旋CT图像常见伪影的分析和处理 被引量:11

Analysis and Processing of Artifacts on Helical CT Images
在线阅读 下载PDF
导出
摘要 目的研究螺旋CT成像中常见伪影的产生原因与处理方法,以减少螺旋CT检查时产生的伪影,提高图像质量,为临床提供良好的影像资料。方法收集本院156位患者不同类型和不同程度CT图像伪影病例,进行系统性分析,其中头颈部扫描61例,胸部扫描37例,腹部扫描33例,盆腔扫描15例,其它部位10例。结果针对螺旋CT常见伪影产生的机理和成因进行了分析,给出了避免、抑制以及消除基于物理结构、基于患者本身、基于扫描系统和基于重建方法四类不同图像伪影的处理方法。结论螺旋CT成像中伪影产生受多种因素影响,利用各种技术有效避免、抑制或消除图像伪影,对优化图像质量和提高临床医学诊断质量和效率具有重要意义。 Objective By studing the common causes and treatment methods of spiral CT imaging artifacts,in order to reduce the generated artifacts,improve image quality,provide a good image for the clinical data. Methods CT images with different types and different degrees artifacts were collected and analyzed from 156 radiological examination cases in our hospital. Among them,61 cases received head and neck scan,S7 cases chest scan,33 cases abdomen scan, 15 cases pelvic scan, 10 cases other location scan. Results For the common artifacts on helical CT images, their generating mechanism and reasons are analyzed. For the physical-based, patient-based, scanner-based artifacts and reconstruction-based artifacts, the image processing solutions are proposed in detail. Conclusion Understanding the various artifacts of the mechanism, and using a variety of effective techniques to avoid, suppress and eliminate image artifacts are very important for optimizing the CT image and improving its quality and efficiency in the application of clinical diagnosis.
出处 《中南医学科学杂志》 CAS 2011年第6期692-694,共3页 Medical Science Journal of Central South China
关键词 计算机断层扫描 螺旋CT 伪影 图像处理 computed tomography helical CT artifact image processing
  • 相关文献

参考文献5

二级参考文献18

  • 1李新华,李强,张海宁,崔文国,李荣环,张振.医用CT的质量控制和保证[J].医疗卫生装备,2004,25(7):39-39. 被引量:16
  • 2黄晓明,吴向阳,张超群.PET/CT图像融合技术[J].医疗卫生装备,2004,25(9):41-42. 被引量:6
  • 3杨克柽,叶育盛,余晓锷,林意群.三维CT成像的数学模型[J].CT理论与应用研究(中英文),2006,15(1):37-41. 被引量:2
  • 4周卫兵,欧陕兴,杨克柽.螺旋CT梯状伪影的数学模型及实验研究[J].医疗设备信息,2006,21(4):5-8. 被引量:7
  • 5Wesolowski JR,Lev MH.CT:history,technology,and clinical aspects.Semin Ultrasound CT MR,2005,26 (6):376-379.
  • 6PeiJan PL,Thomas JB,Caridad B,et al.Specificationandacceptance testing of computed tomography scanners.AAPM Report No.39,1993.
  • 7Chen GT,Kung JH,Beaudette KP.Artifacts in computed tomography scanning of moving objects.Semin Radiat Oncol,2004,14 (1):19-26.
  • 8Kalender WA,Hebel R,Ebersberger J.Reduction of CT artifacts caused bymetallicimplants.Radiology,1987,164 (2):576-577.
  • 9Fleischmann D.Rubin GD.Paik DS.Yen SY.Hilfiker PR.Beaulieu CF.Napel S.Stair-step artifacts with single versus multiple detector-row helical CT.Radiology,2000,216 (1):185-196.
  • 10Heuscher DJ,Vembar M.Reduced partial volume artifacts using spiral computed tomography and integrating interpolator[J].Med.Phy,1999,26 (2):276-286.

共引文献22

同被引文献77

  • 1李新华,李强,张海宁,崔文国,李荣环,张振.医用CT的质量控制和保证[J].医疗卫生装备,2004,25(7):39-39. 被引量:16
  • 2侯宗来,谢良军,张云霈.多功能头托在CT检查中的应用[J].中国中西医结合影像学杂志,2005,3(3):237-237. 被引量:1
  • 3杨克柽,王余锋.CT图像伪影[J].医疗装备,2006,19(5):16-17. 被引量:10
  • 4张春玲,亓恒涛.多层螺旋CT图像伪影的探讨[J].泰山医学院学报,2006,27(6):550-552. 被引量:9
  • 5张春玲,亓恒涛.多层螺旋CT图像伪影的探讨[J].泰山医学院学报,2006,27(5):425-427. 被引量:2
  • 6Yazdi M,Beaulieu L.Artifacts in spiral X ray CT scanners.problems and solutions[J]. International Journal of Biological Medical Sciences, 2009,4(3). 135-139.
  • 7Hsieh J.Image artffacts.appearances causes and corrections. Computed tomography.principles design artifacts and recent advances[M]. Bellingham Wash: SPIE Press, 2003: 167-240.
  • 8曹明干,叶昌印,杨守营,陈幸福,徐嘉彬.飞利浦Brilliance 16 排CT 故障原因分析及其保养[A].2011 年浙江省医学会医学工程学分会第九届学术年会论文汇编[C].2011,30(11):79-82.
  • 9蔡永辉,吴新来.我院GE Lightspeed Ultra CT 生命周期总结[A].2011 年浙江省医学会医学工程学分会第九届学术年会论文汇编[C].2011,12(21):68-71.
  • 10Alvarez R E, Macovski A. Energy-selective reconstruc- tions in X-ray computerised tomography [ J ]. Physics in Medicine and Biology, 1976, 21 (5) : 755.