期刊文献+

应用UF-100全自动尿沉渣分析仪检测浆膜腔积液的作用和意义 被引量:8

Application value of UF-100 automated urinalysis analyzer in serous fluid detection
在线阅读 下载PDF
导出
摘要 目的:探讨UF-100尿液分析仪检测浆膜腔积液的临床价值和意义。方法:运用传统手工镜检法和UF-100全自动尿液分析仪法同时检测患者浆膜腔积液,并比较两种方法对白细胞的检测结果,比较使用UF-100全自动尿液分析仪法检测不同性质浆膜腔积液标本白细胞、间皮细胞和细菌数。结果:两种方法中白细胞数(WBC)间存在较高的相关性(P>0.05),漏出性与渗出性浆膜腔积液间白细胞数、间皮细胞、细菌数均存在显著性差异(P<0.05),非化脓性与化脓性细菌感染性积液中WBC数、细菌数差异均有显著性(P<0.05),间皮细胞数差异无显著性(P>0.05)。结论:用UF-100尿液分析仪检测胸腹水标本简单、易行,检测结果对相关疾病有更好的临床价值。 Objective To discuss the clinical application value of UF-10O0 automated urinalysis analyzerr (called UF-100 in short) in examination of white blood cell (WBC) count in serous fluid. Methods Anticoagulant samples were directly analyzed by UF-100 and manual microscopic method (called manual method in short), in order to compare the results of the serous fluid samples among WBC epithelia and bacteria between these two methods. Results There was significant relativity the between two methods. There was remarkable differences in the result of WBC and bacteria (P 〈 0.05) analyzed by these two methods, but epithelia showed significant difference (P 〉 0.05). Conclusions Using UF- 100 in examination of WBC count in serous fluid was simple and applicable, and there was clinical application value. It may replace the complex traditional routine examination of ascites or hydrothorax.
出处 《实用医学杂志》 CAS 北大核心 2010年第2期308-310,共3页 The Journal of Practical Medicine
关键词 白细胞 全自动尿液分析仪 浆膜腔积液 Leukocyte Urine analysis instrument Serous fluid
  • 相关文献

参考文献7

二级参考文献17

共引文献205

同被引文献46

  • 1魏明,甘文.UF-100全自动尿沉渣分析仪在胸腹水检查中的应用[J].江西医学院学报,2004,44(5):107-107. 被引量:16
  • 2白垚,程大林,张济.自动化仪在胸腹腔积液细胞计数中应用的探讨[J].重庆医科大学学报,2005,30(6):859-861. 被引量:13
  • 3杨海芸,戈之铮.粪便隐血试验在结直肠癌早期筛查中的研究现状[J].胃肠病学,2006,11(6):376-378. 被引量:9
  • 4王升,张红升.无菌体液细菌培养的新方法探讨[J].实用医技杂志,2007,14(19):2668-2669. 被引量:2
  • 5周新,府伟灵.临床生物化学与检验[M].4版.北京:人民卫生出版社,2008:50-51.
  • 6Smith R A,Cokkinides V,Eyre H J.America Cancer Society guidelines for the early detection of cancer[J].CA Cancer J Clin,2006,56(3):11-25.
  • 7Levi Z,Hazazi R,Rozen P,et al.A quantitative immunochemical fecal occult blood test is more efficient for detecting significant colorectal neoplasia than a sensitive guaiac testr[J].Alinment Pharmacol Ther,2006,23(5):1359-1364.
  • 8Scales C D Jr,Fein S,Muir A J,et al.Clinical utilization of digital rectal examination and fecal occult blood testing upon hospital admission r[J].J Clin Castroenterol,2006,40(8):913-918.
  • 9Vilkin A,Rozen P,Levi Z,et al.Perfomance characteristics and evaluation of an automated-developed and quantitative,immunochemical,fecal occult blood screening testr[J].Am J Castroenterol,2005,100(1):2519-2525.
  • 10de Jonge R, Brouwer R, de Graaf M T, et al. Evaluation ofthe new body fluid mode on the Sysmex XE - 5000 for count- ing leukocytes and erythrocytes in cerebrospinal fluid and oth- er body fluids[J]. Clin Chem Lab Me, 2010, 48(5) ; 665.