REXEED^(TM)系列透析器在高通量透析中的溶质清除特性 被引量:1

Solute clearance characteristics of REXEED^(TM) series dialyzer during high-flux dialysis
在线阅读 下载PDF
导出
摘要 目的研究REXEEDTM系列透析器的溶质清除特性,讨论护理特点。方法规律性血液透析患者18例,采用随机化交叉和3×3拉丁方实验设计,每例患者分别运用REXEEDTM-15AC、15UC和对照透析器APS-15U进行血液透析治疗。于透析器血液入、出口处采血测定尿素氮、肌酐、磷、β2-微球蛋白,计算溶质清除率;在每使用一种透析器的第3次透析前后,采血测定患者血尿素氮和肌酐水平,计算下降比率。严密观察患者生命体征,监测透析机的静脉压和跨膜压数值。结果REXEEDTM-15AC和REXEEDTM-15UC透析器的尿素氮清除率显著大于APS-15U透析器(P<0.05);REXEEDTM-15AC透析器肌酐清除率显著大于APS-15U透析器(P<0.05)。同一患者使用不同透析器的血尿素氮下降比率比较差异无统计学意义(均>65%)。血液透析过程中,患者生命体征稳定,无不良反应发生;实验室安全性指标监测未见异常。结论REXEEDTM系列透析器临床使用安全、有效。透析护理过程中,必须加强观察并重视患者的不适主诉;对于超滤量少的患者,可在适当补充液体的同时增加超滤量,以提高跨膜压,防止反超。 Objective To evaluate the solute clearance characteristics of REXEEDTM series dialyzers during high-flux dialysis, and explore the care characteristics. Methods A randomized crossover study of 3 × 3 Latin square was designed based on different dialyzers. Eighteen patients with regular henlodialysis underwent dialysis with REXEED^TM-15AC dialyzer, REXEEDTM-15UC dialyzer and controlled APS-15U dialyzer, respectively. Blood samples were obtained from the blood flow entrance and exit of dialyzers, levels of urea nitrogen, creatinine, phosphate and 132-microglobulin were detected, and solute clearance rates were calculated. Before and after the third dialysis with each dialyzer, blood samples were obtained to measure the levels of urea nitrogen and creatinine, and the rates of decrease were calculated. The vital signs of each patient were intensively observed, and the venous pressure and transmembrane pressure were monitored from the dialyzers. Results The urea nitrogen clearance rates of REXEED^TM-15AC dialyzer and REXEED^TM-15UC dialyzer were significantly higher than that of APS-15U dialyzer ( P 〈 0. 05). The creatinine clearance rate of REXEED^TM-15AC dialyzer was significantly higher than that of APS-15U dialyzer(P 〈 0.05). There was no significant difference in the rate of decrease in blood urea nitrogen among different dialyzers of the same patient( 〉 65% for all patients). The vital signs were stable with no adverse events during dialysis, and there was no abnormal findings in laboratory security parameters. Conclusion REXEED^TM series dialyzers are effective and safe for clinical application. Great importance should be attached to the complaints from patients during dialysis. For those with less ultrafiltration, fluid as well as uhrafihration should be supplemented to increase the transmembrane pressure.
出处 《上海交通大学学报(医学版)》 CAS CSCD 北大核心 2009年第7期858-861,共4页 Journal of Shanghai Jiao tong University:Medical Science
关键词 高通量透析 透析器 清除率 护理 high-flux hemodialysis dialyzer clearance care
  • 相关文献

参考文献9

  • 1Jacobs C, Kjellstrand CM, Koch KM, et al. Replacement of renal function by dialysis[ M ]. Dordrecht : Kluwer Academic, 1996 : 2 - 102.
  • 2Brendolan A, Nalesso F, Fortunato A, et al, Dialytic performance evaluation of RexeedTM: A new polysulfone-based dialyzer with improved flow distributions [ J ]. Int J Artif Organs, 2005, 28 (10) : 966 -975.
  • 3Fukuda M, Miyazaki M, Uezumi S, et al, Design and assessment ta the new APS dialyzer (APS-SA series) [J].J Artifs Organs, 2006, 9(3): 192-198.
  • 4Yamamoto K, Matsukawa H, Yakushiji T, et al, Technical evalua- tion of dialysate flow in a newly designed dialyzer [ J ]. ASAIO J, 2007,53(1) : 36 -40.
  • 5龚德华.对血液透析尿素动力学模型及透析充分性的一些新认识[J].肾脏病与透析肾移植杂志,2001,10(3):278-282. 被引量:16
  • 6龚德华,季大玺,陈惠萍,谢红浪,徐斌,刘芸,黎磊石.高通量维持性血液透析患者甲状旁腺素水平及骨代谢的变化[J].肾脏病与透析肾移植杂志,2001,10(4):317-321. 被引量:85
  • 7Leto E, Bilal F, Osmic I. Efficiency of high-flow dialyzers in removal of beta-2-mcroglobulin [J].Med Arh, 2001, 55 (4) : 225 - 226.
  • 8Gerdemann A, Lemke HD, Nothdurft A, et al. Low-molecular but not high-molecular advanced glycation end products (AGEs) are removed by high-flux dialysis [ J]. Clin Nephrol, 2000, 54 (4) : 276 - 283.
  • 9Lang SM, Bergner A, Topfer M, et al. Preservation of residual renal function in dialysis patient: effects of dialysis-technique-related factors[ J]. Petit Dial Int, 2001, 21 ( 1 ) : 52 - 57.

二级参考文献5

共引文献97

同被引文献6

引证文献1