评价电子射野影像系统验证鼻咽肿瘤放疗摆位误差时对患者剂量的影响 被引量:4

Evaluate the Dose Affection on Patient in the Process of Validating Set-up Error by Electronic Portal Imaging Device (EPID) for Nasopharyngeal Carcinoma (NPC)
在线阅读 下载PDF
导出
摘要 目的:分析利用电子射野影像系统(EPID)纠正鼻咽癌调强放射治疗(IMRT)的摆位误差时,对患者剂量的影响。方法:选取IMRT的鼻咽癌患者30例,每拍一次电子射野影像片(EPI),拍摄正位片(机架角为0°)和侧位片(机架角为90°)共计2张。拍摄正、侧位片时射野面积一般取10cm×10cm,机器跳数为3MU-4MU。利用荷兰核通公司的Masterplan三维治疗计划系统(Version1.5)进行剂量计算统计。结果:每拍一次EPI,机器跳数为3MU,当眼晶体在射野范围之内时,射线是能量为6MV的光子束,左、右眼晶体平均剂量为4.8cGy左右;射线是能量为8MV的光子束,左、右眼晶体平均剂量为4.6cGy左右。脊髓和脑干受照部分的平均剂量为4.7cGy左右。如果拍一次EPI机器跳数为4MU,器官受照部分的平均剂量要相应增加33%。结论:建议在鼻咽癌放疗前,都于EPID下进行实时摆位误差纠正,而在设计鼻咽癌治疗计划时,需将这部分剂量考虑到计划中,以避免剂量误差。更好的方法是在治疗计划中设计拍EPI的射野时,用准直器、MLC保护眼晶体、脊髓和脑干等危及器官。 Objective: To analyse the extra dose affection on patient in the process of validating set-up error by electronic portal imaging device (EPID)for nasopharyngeal carcinoma (NPC) intensity modulated radiation therapy (IMRT). Methods: Thirty patients were enrolled in study.Electronic portal images (EPI) were aquired with field size 10 cm×10 cm, 3 MU-4 MU,the anterior and lateral EPI were included in the radiation therapy course. And then,analyzing the extra dose affection on patient by Oncentra Materplan (Version 1.5SP1,Nuclectron). Results: When the electronic portal images with 3 MU, 6 MV X-ray were done,the average exposure dose of the left or right lens is about 4.8 cGy. And 8 MV X-ray, about 4.6 cGy on left or right lens, 4.7 cGy on spinal cord and brain stem. If the EPI is done with 4 MU, the average dose of organ at risk(OAR)will increase 33%. Conclusions: So as to validate the set-up error, we suggest that electronic portal image should be required before the NPC radiation treatment. Besides, In designing the treatment plan phase, we should consider the extra dose affection on patient in the EPI process. The best method is that we protect the organ at risk (lens, spinal cord, brain stem, etc.) by jaws or multileaf collimator in the course of EPI.
出处 《中国医学物理学杂志》 CSCD 2008年第3期627-629,共3页 Chinese Journal of Medical Physics
关键词 鼻咽肿瘤 调强放射疗法 电子射野影像系统 nasopharyngeal neoplasms intensity modulated radiation therapy(IMRT) electronic portal imaging device (EPID)
  • 相关文献

参考文献6

二级参考文献42

  • 1林承光,邓小武,黄峻,温志祥,肖达宜,林览峰.鼻咽癌适形放疗实施过程精确度和重复性研究[J].肿瘤学杂志,2004,10(4):208-210. 被引量:32
  • 2郑步宏,潘建基,徐鹭英,陈传本,张瑜,张秀春.垫内置标记点法测量鼻咽癌适形放疗摆位误差的研究[J].中华放射肿瘤学杂志,2005,14(3):218-221. 被引量:19
  • 3翟振宇,何侠,魏青,张宜勤.鼻咽癌调强放射治疗的剂量学特点[J].肿瘤学杂志,2005,11(5):337-340. 被引量:21
  • 4何侠,翟振宇,纪红,张宜勤.初治鼻咽癌调强适形放疗近期临床观察[J].肿瘤防治研究,2006,33(4):272-274. 被引量:31
  • 5胡逸民.治疗计划设计的物理原理和生物学基础[A].见:胡逸民主编.肿瘤放射物理学[C].北京:原子能出版社,1999.328-331.
  • 6[2]Kinxie JJ, Hanks GE, Maclean, CJ et al. Patterns of care study: Hodgkin's Disease relapse rates and adequacy of portals[J]. Cancer, 1983, 52:2223-2226.
  • 7Kam MK, Chau RM, Suen J, et al. Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalation [J]. Int J Radiat Oncol Biol Phys, 2003, 56(1):145-157.
  • 8van Lin EN, van der Vight L, Huizenga H, et al. Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customised head and neck support[J]. Radiother Oncol, 2003,68(2):137-148.
  • 9Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability[J]. Int J Radiat Oncol Biol Phys, 1999,43:905-919.
  • 10Schallenkamp JM, Herman MG, Kruse JJ, et al. Prostate position relative to pelvic bony anatomy based on intraprostatic gold markers and electronic portal imaging [J].Int J Radiat Oncol Biol Phys, 2003, 63:800-811.

共引文献78

同被引文献30

  • 1林承光,邓小武,黄峻,温志祥,肖达宜,林览峰.鼻咽癌适形放疗实施过程精确度和重复性研究[J].肿瘤学杂志,2004,10(4):208-210. 被引量:32
  • 2沈捷,张福泉,邱杰,胡克,张明杰.应用EPID对放射治疗摆位误差的研究[J].中华放射医学与防护杂志,2006,26(3):308-308. 被引量:39
  • 3彭振军,伍钢,王晔,徐宁,周妍,冯敢生.放疗射野影像验证中EPID不同摄影剂量的影像效果评价[J].华中科技大学学报(医学版),2007,36(3):395-396. 被引量:3
  • 4胡逸明,杨定宇.肿瘤放射治疗技术[M].北京:北京医科大学·中国协和医科大学联合出版社,2007:167.
  • 5陈明伟,邓小武,黄劭敏,陈利,康德华.非晶硅平板电子射野影像用于放射治疗剂量学质量控制检验的应用[J].癌症,2007,26(11):1272-1275. 被引量:12
  • 6HuYM,DaiJR,HuB.颅脑肿瘤中立体定向放射治疗靶区位置准确性分析[R].立体定向放射治疗国际讨论会1997忠南国立大学医学院韩国,1993.
  • 7Sang-JunePark,DanIonascu,FredHacker,HarveyMamon,andRossBerbeco立体定向放射治疗中使用EPID图像进行金球的自动探测和三维重建[J].医学物理,36(10)102009.
  • 8Nederveen,J.Lagendijk,andP.Hofman,使自动实时MV级x线位置验证中使用标志物提取核的金球探测[J].放射生物和物理杂志47期1435-1442页2000.
  • 9Nederveen,J.Lagendijk,andP.Hofman,使用非晶硅平板成像设备的标志物自动探测可行性[J].放射生物和物理杂志46期第1219-1230页(2001).
  • 10W.Mao,R.D.Wiersma,andL.Xing,利用MV和KV级成像系统的快速治疗中标志物追踪算法[J].医学物理35期,1942-1949(2008).

引证文献4

二级引证文献30

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部