期刊文献+

离子色谱法测定浓碱中痕量阴离子 被引量:7

Ion Chromatographic Determination of Trace Anions in High Alkali Solutions
在线阅读 下载PDF
导出
摘要 采用离子交换树脂处理以除去干扰样品的氢氧根离子,很好地消除了基线效应。以Dionex AS9-HC作为分离柱,9mmol/L碳酸钠溶液作为淋洗液,抑制电导检测。该方法的结果令人满意,基线平直,检出限低,回收率在96%-104%之间,对于测定浓碱中的阴离子是一种很有效的方法。 Anion exchange resin was used to remove hydroxyl ions, which effectively eliminate the domino effect of baseline. Dionex IonPac AS9-HC column was used for separation and 9 mmol/L sodium carbonate solution was used as the eluent to suppress conductivity detection. The results were satisfactory and the detection limits of these anions were low. Recoveries from spiked samples were between 96% and 104% . Compared with other methods, it is an efficient and simple method for the determination of anions in sodium hydroxide. This method can be used for determination of anions in high alkali solution and solid alkalis.
出处 《分析测试学报》 CAS CSCD 北大核心 2006年第2期112-114,共3页 Journal of Instrumental Analysis
基金 国家自然科学基金资助项目(20375035) 浙江省自然科学基金资助项目(Z404105) 教育部及浙江省留学回国人员科研启动基金 分析测试基金资助项目
关键词 离子色谱 浓碱 阴离子 Ion chromatography High alkali solution Anion
  • 相关文献

参考文献4

  • 1SERGEEV G M,LUKUTTSOV A A,ZORIN A D.Determination of trace anions in organic solvents[J].Anal J Chem(Russia),1992,49(4):391 -395.
  • 2牟世芬 刘克纳.离子色谱方法及应用[M].北京:化学工业出版社,2001.228-248.
  • 3SIRIRAKS A,KINGSTON H M,RIVIELLO J M.Chelation ion chromatography as a method for trace elemental analysis in complex environmental and biological samples[J].Anal Chem,1990,62:1185-1193.
  • 4WANG K F,LEI Y,EITEL A,et al.Ion chromatographic analysis of anions in ammonium hydroxide,hydrofluoric acid,and slurries,used in semiconductor processing[J].J Chromatogr,A.,2002,(956):109 -120.