直接眶压检测方法及眶压正常值的研究 被引量:2

Direct measurement of orbital tension in normal orbits
原文传递
导出
摘要 目的研究直接眶压的测量方法、正常值及影响因素。方法使用自行设计组装的直接眶压计对40例40只正常眼眶进行测量,同时测量眼压、眼球突出度与间接眶压。结果我国正常人的直接眶压值为(49±12)mmHg。直接眶压与眼压有正相关关系(r=056),与年龄有负相关关系(r=-040)。结论该直接眶压测量仪操作简便、创伤性小,结果准确稳定、能客观反映眼眶内的真实压力。 Objective To measure orbital tension in a normal orbit group and investigate some factors involved in.Methods Orbital tensions were directly measured by a designed orbitonometer for 40 normal orbits and in the meantime,the intraocular pressures and the amounts of exophthalmos were recorded.Results Orbital tension in the normal orbit group was (4 9±1 2)mmHg and a positive correlation was showed between the intraocular pressure and orbital tension(r 2=0 31),whereas the age was negatively correlated with the orbital tension(r 2=0 16).Conclusions Our orbitonometer may safely and accurately measured orbital tension.
出处 《中国实用眼科杂志》 CSCD 北大核心 2004年第11期874-876,共3页 Chinese Journal of Practical Ophthalmology
基金 上海市卫生系统百名跨世纪优秀学科带头人培养计划(98BR005) 上海市青年科技启明星计划(00QNH1460)
关键词 正常值 眼压 直接 正常眼 创伤性 眼眶内 正常人 结论 准确 中国 Orbit Orbital tension Orbitonometer Intraocular pressure
  • 相关文献

参考文献12

  • 1Kratky V, Hurwitz JJ, Avram DR.Orbital compartment syndrome. Direct measurement of orbital tissue pressure: 1. Technique. Can J Ophthalmol, 1990, 25:293~297
  • 2McGowan HD, Hurwitz JJ, Gentles W.Orbitotonography, the dynamic assessment of orbital tension: 1 .Results in subjects without known orbital disease. Can J Ophthalmol, 1984, 19: 122~125
  • 3Hurwitz JJ, McGowan HD, Gentles W, et al. Orbitotonography,the dynamic assessment of orbital tension: 1.Results in patients with orbital disease. Can J Ophthalmol, 1988, 23:308~310
  • 4Frueh BR, Musch DC, Gril R, et al. Orbital compliance in Graves' eye disease. Ophthalmology, 1985, 92:657~665
  • 5Stanley RJ, McCaffrey TV, Offord KP, et al. Space- occupying orbital lesions: can critical increases in intraorbital pressure be predicted clinically. Laryngoscope, 1989, 99:19~22
  • 6Riemann CD, Foster JA, Kosmorsky GS. Direct orbital manometry in healthy patients. Ophthal Plast Reconstr Surg, 1999, 15:121~125
  • 7Riemann CD, Foster JA, Kosmorsky GS. Direct orbital manometry in patients with thyroid- associated orbitopathy. Ophthalmology,1999, 106:1296~1302
  • 8刘玉珉,吕昊,魏光杰,李钟实.直接接触式眶压计及正常人眶压的研究[J].眼科新进展,2000,20(4):280-281. 被引量:9
  • 9Forrest CR, Khairallah E, Kuzon WM, et al. Intraocular and intraorbital compartment pressure changes following orbital bone grafting: a clinical and laboratory study. Plast Reconstr Surg,1999, 104:48~54
  • 10Hayreh SS, Kolder HJ, Weingeist TA.Central retinal artery occlusion and retinal tolerance time. Ophthalmology, 1980, 87:75 ~78