乳腺影像学专用PET探测器模块设计

Design of LYSO Detector for Positron Emission Tomography Breast Imaging
在线阅读 下载PDF
导出
摘要 为进一步提升乳腺影像学专用PET系统的空间位置分辨率,基于LYSO晶体与位置灵敏光电倍增管耦合,设计了一种专用LYSO-PET探测器模块.探测模块由12×12根1.5 mmx1.5 mmx10 mm LYSO晶体条阵列与PS-PMT耦合组成,通过PXI 8501采集系统,获取入射粒子的2D泛场图像、能量信息及位置信息.该探测模块平均位置分辨率可达0.22 mm(本征)和0.55 mm(^(22)Na探测条件下);平均能量分辨率为22.4%(^(22)Na探测条件下).双探头PET系统FOV中心^(22N)a点源轴向空间分辨率可达1.13 mm.探测器模块已基本满足位置分辨率要求,为实现亚毫米乳腺影像学专用PET系统提供了硬件支持. Developed a LYSO scintillation detector module based on PS-PMT for Positron emission tomography,to improve the positional resolution.The module consists of a 12×12 array of 1.5 mm×1.5 mm×10 mm(pitch 1.6 mm)LYSO elements.The pulse signal is read by PXI8501 to obtain 2D panfield image,energy and position information.The module has an average position resolution of 0.22 mm(intrinsic)and 0.55 mm(^(22)Na),an average energy resolution of 22.4%with^(22)Na.The dual-detector simplified PET system has a spatial resolution at the center of the FOV of 1.13(axial)with^(22)Na.The detector module can meet the requirements of breast imaging-specific PET,which will have submillimeter positional resolution performance.
作者 范亚明 尹永智 FAN Ya-ming;YIN Yong-zhi(The third the people's hospital of Bengbu,,Bengbu 230000,China;School of Nuclear Science and Technology,Lanzhou University,Lanzhou 730000,China)
出处 《核电子学与探测技术》 CAS 北大核心 2022年第5期874-878,共5页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金(11875156)资助。
关键词 硅酸钇镥闪烁晶体(LYSO):乳腺PET 位置分辨率:能量分辨率 lutetium yttrium orthosilicate bregt PET position resolution energy resolution
  • 相关文献

参考文献4

二级参考文献22

  • 1高道利,胡永伟,王文婉,陈范良,潘雷达,袁亚,郁领娣,钱锋.乳房自我检查对降低女性乳腺癌死亡率干预效果的评估[J].中华流行病学杂志,2006,27(11):985-990. 被引量:14
  • 2H Zaidi, C Thompson. Evolution and Developments in Instrumentation for Positron Emission Mammography [J]. Pet Clinics, 2009, 4(4) :317 -327.
  • 3C. J. Thompson. Feasibility study for positron emis- sion mammography [J]. MEDICAL PHYSICS, 1994,21 (4) :529 -538.
  • 4MC Abreu, JD Aguiar, F'G Almeida, et al. Design and Evaluation of the Clear - PEM Scanner for Posi- tron Emission Mammography [K]. IEEE Transactions on Nuclear Science, 2006, 53 (1) :71 - 77.
  • 5LI Lin, GU Xiao - yue, LI Dao-wu, et al. Perform- ance Evaluation and Initial Clinical Test of the Posi- tron Emission Mammography System (PEMi) [J]. IEEE Transactions on Nuclear Science, 2015, 62 (5):1 -9.
  • 6Guerra A D, Belcari N, Bisogni M G, et al. Silicon Photomultipliers (SiPM) as novel photodetectors for PET [J]. Nuclear Instruments & Methods in Physics Research, 2011, 648(17) :$232 $235.
  • 7V Popov, S Majewski, AG Weisenberger. Readout e- lectronics for multi - anode photomultiplier tubes with pad matrix anode layout [J]. IEEE Nuclear ScienceSymposium Conference Record, 2003, 567( 1 ):2156 -2159 Vol. 3.
  • 8Neves J A. The Clear PEM breast imaging scanner [J]. Nuclear Instruments & Methods in Physics Re- search A, 2011, 628(1):444-447.
  • 9Moadel R M. Breast cancer imaging devices [J]. Seminars in Nuclear Medicine, 2011, 41 ( 3 ) : 229 241.
  • 10Li L, Gu X, Li D, et al. Ability of the Positron E- mission Mammography system, PEMi, in detection of millimeter- sized lesions [C]// Nuclear Science Symposium and Medical Imaging Conference (NSS/ MIC), 2013 IEEE. IEEE, 2013:1 -7.

共引文献1318

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部