某区2013年放射诊疗设备机房周围环境防护检测结果分析 被引量:1

Analysis on test of a military medical radiation equipment room protection in 2013
原文传递
导出
摘要 目的:为了掌握某军区33家医疗卫生单位2013年放射诊疗设备机房周围环境辐射防护状况。方法:现场检测、数据分析并依据相应的国家标准进行评价。结果:普通X线机房各检测项目中不合格的项目主要是防护门、防护墙和防护窗,CT机房不合格的项目主要是防护门和楼面,核医学设备机房不合格的项目主要是休息室防护门及工作场所的表面污染。三种类型设备总的防护检测合格率由低到高的顺序是:核医学、放射诊断和放射治疗。其他设备机房防护检测合格率均为100%。结论:应重视改建机房放射防护的设计和施工,加强对CT机房防护门及楼面的防护,加强对PET-CT工作场所注射过放射性药物患者的管理,核医学工作场所必须随时和定期进行放射性表面污染监测。 Objective:To grasp the environmental radiation levels around radiation equipment of 33 military health units in2013.Methods Methods:The methods were field test,data analysis and evaluation according to the corresponding national standards.Re Results:The unqualified projects of ordinary X-ray were mainly protective doors,walls and windows,CT unqualified projects were doors and floors,nuclear medical equipment unqualified projects were anteroom doors and surface contaminations.Three kinds of equipment qualified rates from high to low order was:nuclear medicine,diagnostic radiology and radiation therapy.In addition to the ordinary X-ray,CT and PET-CT device,detection qualified rate was 100%.Conclusion:Attention should be paid to the design and construction of the reconstructed room radiological protection,strengthen the protection of CT room doors and floors,strengthen the management of patients with radioactive drugs on PET-CT workplace,and readily and regularly monitor surface contamination must be done in nuclear medical workplaces.
作者 赵国良 程远泽 李勇 朱乐明 金慧英 杨龙 ZHAO Guo-liang;CHENG Yuan-ze;LI Yong;ZHU Le-ming;JIN Hui-ying;YANG Long(Medical Protection Institute,Center for Disease Control and Prevention,PLA Nanjing Military Region,Jiangsu Nanjing,210002,China)
出处 《灾害医学与救援(电子版)》 2014年第2期90-93,共4页 Disaster Medicine and Rescue(Electronic Edition)
关键词 放射防护 监督监测 周围剂量当量 去污 radiation protection supervision of ambient ambient dose equivalent rate decontamination
  • 相关文献

参考文献6

二级参考文献42

  • 1邓大平,李全太.医用电离辐射防护中存在的一些问题[J].中华放射医学与防护杂志,2005,25(4):407-408. 被引量:12
  • 2GBZ166-2005,职业性皮肤放射性污染个人监测规范[S].
  • 3Brenner DJ, Hall EJ. Computed Tomography - An Increasing Source of Radiation Exposure [ J ]. N Engl J Med,2007,357 ( 22 ) : 2277 -2284.
  • 4Kalra MK,Maher MM ,Toth TL,et al. Strategies for CT Radiation Dose Optimizationl [ J ]. Radiology,2004,230:619-628.
  • 5Slovis TL. CT and Computed Radiography:The Pictures Are Great, but Is the Radiation Dose Greater than Required? [ J] AJR, 2002,179 ( 1 ) :39-41.
  • 6McNitt-Gray MF. Radiation Dose in CT[ J]. Radiographics,2002,22 : 1541-1553.
  • 7Goldman LW. Principles of CT:Radiation Dose and Image Quality[J]. Journal of Nuclear Medicine Technology,2007,35 (4) :213-225.
  • 8Mayo JR, Aldrich J, Muller NL,et al. Radiation Exposure at Chest CT: A Statement of the Fleischner Society [ J ]. Radiology, 2003,228 : 15- 21.
  • 9Jacobi W. The Concept of Effective Dose: A Proposal for the Combination of Organ Doses [ J ]. J Radiat Environ Biophys, 1975, 12 : 101- 109.
  • 10McCollough C and Schueler B. Educational Treatise : Calculation of Effective Dose[J]. Med Phys,2000, 27(5) :828-837.

共引文献20

同被引文献11

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部